Figure 8: Position 'D' measures diffuse reflectance only. Position 'S' measures total
reflectance by including the specular component.
The sphere has five ports:
1. Sample beam transmission port (19 x 17 mm)
2. Sample measurement port (16 mm diameter)
3. Reference beam port (11 x 13 mm)
4. NIR detector port (11 mm diameter)
5. PMT port (oval, 30 mm long x 14 mm wide)
The PMT, PbS and InGaAs detectors are mounted on top of the sphere, which is
alternately illuminated by sample and reference beams.
Beam image dimensions:
transmission port 11.04 mm (h) x 13.44 mm (w)
reflectance port 12.50 mm (h) x 7.11 mm (w)
Figure 9: The beam through the accessory
The reference disks supplied with the accessory are 6 mm deep and the powder is
added to a density of 1 g/cm
3
. This is the optimum thickness and density of PTFE, to
ensure maximum reflectivity(3) and therefore the best absolute accuracy.
Specifications
+ 85 volts DC (Cary
4000/5000/6000i)
DRA 900 DRA 1800 DRA 2500
Cary 4000
200 nm to
900 nm *
200 nm to
900 nm
200 nm to
900 nm
Cary 5000
200 nm to
900 nm *
200 nm to 1800
nm *
200 nm to
2500 nm *
Wavelength range
Cary 6000i
200 nm to
900 nm *
200 nm to
1800 nm *
200 nm to 1800
nm