Figure 29. The mask fits onto the double aperture base. This figure shows only one
aperture is open.
7. Click the Windows Start button, then Programs, then Cary WinUV, and then
Align.
8. Set the SBW to 2.00.
9. Close the sample compartment lid.
10. Tick the Zero Order check box, click the Apply button to drive the instrument
to zero.
11. Open the sample compartment lid and adjust the height adjustment screw on
the side of the polarizer mount until the image is centred over the apertures.
Hot tip
Holding a piece of white paper in front of the apertures will made the beam
image easier to see. You can use the aperture mask to draw the apertures on the
paper before mounting the mask.
Figure 30. The image should be centred over the apertures.
12. Close the sample compartment lid and go to the wavelength that you are going
to use for the measurement. Take note of the % transmission reading.
13. Open the sample compartment lid and rotate the aperture mask so that only the
lower aperture is open.