Figure 16. The sample mount ledge.
13. Click the Start button. The Save As dialog box will appear, allowing the
method to be saved as either a data file or a batch file. If the file is saved as a
batch file, all of the method parameters will be stored with the scan.
This procedure will measure the % reflectance relative to the PTFE reference disk.
Note
To determine the absolute diffuse reflectance of a sample, you need to purchase
an NBS standard and calibrate your PTFE disks. The NBS standards SRM
2021 (a black porcelain enamel tile) or SRM 2019c (a white ceramic tile) are
suitable for this purpose.
To determine position and size of the image
To determine the position and size of the image so that samples can be mounted
correctly, follow this procedure:
Note
Initialize the Cary instrument before placing the DRA into the sample
compartment. After initialization, place the DRA into the compartment.
Do not turn the accessory on (i.e., do not plug the DRA into the instrument).
1. Click the Goto button in the Scan application.
2. Select the Zero order check box and click OK. The spectrophotometer will
drive to 0 nm and select the Vis lamp.
3. Place a piece of white paper in front of the sample clamp. The image size
should now be visible.
NIR reflectance measurements (Cary 5000 and 6000i)
Note
Parameters activated by a radio button that are not specifically mentioned in the
following procedure should be set to 'Off' (e.g. the Signal-to-noise mode radio
button). These parameters will not affect the procedure.
Follow steps 1 to 6
of the transmission measurements procedure then:
Vedere la pagina 34