8. Click the Setup button.
9. From the Setup menu, set the following parameters:
Cary tab
X mode: Mode
N
anometers
X mode: Start 700
X mode: Stop 400
Y mode: Mode % R
Y mode: Y min -5.00
Y mode: Y max 110.00
Scan controls: Ave time (s) 0.100
Scan controls: Data interval (nm) 1.000
Scan controls: Scan rate (nm/min)600.00
Options tab
SBW/Energy: Fixed SBW ON
SBW/Energy: SBW (nm) 2.00
SBW/Energy: Beam mode Double
SBW/Energy: Slit height Reduced
Source: Lamps UV-Vis
Source: Source changeover (nm) 350.0
Baseline tab
Correction Zero/baseline correction
Autostore tab
File Storage: Storage Storage on (Prompt at start)
10. Select the 'Baseline' button from the Scan dialog box. Follow the on-screen
prompts to perform a 100%T baseline scan and a 0%T baseline scan.
Hot Tip
When collecting the 0%T baseline scan do not block the beam, instead remove
the PTFE reference disk from the reflectance port and allow the light to be
trapped by the Extended sample compartment.
11. Once the baseline correction is complete, the Ordinate status display in the top
left corner of the dialog box will show 'Zero baseline' in red text.
12. Clamp the sample over the reflectance port. The sample should be
homogeneous, substantially flat, and large enough to completely cover the
image.
Hot Tip
Use the sample mount ledge for added control over the positioning of your
sample. To use the ledge (see figure 16 ) slide it onto the posts fitted at either
the reflectance or transmission ports.