4. Click the Goto button to open the 'Goto wavelength' dialog box. Enter '801' in
the wavelength field and click OK. The instrument will then drive to 801 nm.
Note
Note the SBW reading in the Status Display dialog box.
5. The SBW reading needs to be at approximately 18. This can be adjusted by
changing the NIR Energy level in the Setup page, under the Independent tab. If
the reading is greater than 18, raise the value of the Energy level field. If the
SBW is under 18, lower the energy level. Once the Energy level has been
adjusted, you will need to click OK in order to note the change to the SBW
reading. Then click the Goto button and again drive the instrument to 801 nm
by entering '801' in the wavelength field and pressing OK. This procedure may
need to be repeated in order to reach the required setting. Please note that each
time you adjust the Energy level reading, you must also drive the instrument to
801 nm.
6. Click the Baseline button in the Scan dialog box. Follow the on-screen
prompts to perform a 100%T baseline scan and a 0%T baseline scan.
Hot Tip
When collecting the 0%T baseline scan do not block the beam, instead remove
the PTFE reference disk from the reflectance port and allow the light to be
trapped by the Extended sample compartment.
7. Once the baseline correction is complete, the Ordinate status display in the top
left corner of the dialog box will show 'Zero baseline' in red text.
8. Clamp the sample over the reflectance port. The sample should be
homogeneous, substantially flat, and large enough to completely cover the
image.
9. Click the Start button. The Save As dialog box will appear, allowing the
method to be saved as either a data file or a batch file. If the file is saved as a
batch file, all of the method parameters will be stored with the scan.
Specular-only reflectance measurements
Note
Parameters activated by a radio button that are not specifically mentioned in the
following procedure should be set to 'Off' (e.g., the Signal-to-noise mode radio
button). These parameters will not affect the procedure.
Follow steps 1 to 6
of the transmission measurements procedure then:
1. From the Setup menu, set the following parameters:
Cary tab
X mode: Mode
N
anometers
X mode: Start 800