The small sample holder is used to measure reflectance of small samples. The powder
cell can be mounted in the small sample holder to enable the reflectance of powder
samples to be measured. The small sample holder is also used to hold the small
sample mount when measuring the reflectance of very small samples.
1. Install the DRA
into the instrument.
2. Screw the posts into position at either the reflectance or transmission port.
3. Slide the small sample holder onto the posts.
4. Secure the sample holder onto with the screws at either side of the attachment.
5. You can now install either the small sample mount
or the powder cell into the
small sample holder.
Note
If you are going to use the powder cell, the small sample holder must be
installed at the reflectance port.
6. You must now check the alignment of the attachment
.
Installing the powder cell
Figure 19. The powder cell. A = lid, B = sample cup, C = funnel.
The powder cell enables you to determine the reflectance of a powder sample.
1. Install the DRA
into the instrument.
2. Install the Small Sample Holder
onto the DRA at the reflectance port.
3. Fit the funnel onto the sample cup.
4. Load your sample into the sample cup.
5. Remove the funnel.
6. Screw on the lid.
7. Insert the powder cell into the small sample holder.